E. Application of User Solution Type

1. User Programmable Temperature Compensation

(Tempco)

This feature allows you to change your Ultrameter III’s temperature

compensation factor to another factor between 0-9.99%/°C (ref.

Temperature Compensation, pg. 56). This feature does not apply to pH

or ORP.

a.

As in Procedure to Select a Solution, pg. 23, select User mode.

b.

With User mode now selected, press

. You may now

adjust a temperature compensation from .00%/°C to 9.99%/°C,

by pressing

or

. See example in Figure 2.

c.

Press

twice to skip calibration adjustment and accept

the new tempco (3 times if in TDS mode). You are now ready to

measure samples with your new temperature compensation factor.

2. Disabling Temperature Compensation

a.

Select User mode (ref. Procedure to Select a Solution, pg. 23).

b.

With “User” selected, press

. If the display does not

show .00%/°C, hold

long enough to bring the tempco to

.00%/°C (see Figure 3).