21

of table salt and tap water (ref. Cleaning Sensors, 2. pH/ORP,

pg. 50). Do not allow pH/ORP sensor to dry out.
I. ORP and Free Chlorine Modes

The Ultrameter III features the ability to measure the activity of oxidizing

or reducing chemicals in solution as ORP mV and the level of residual

sanitizer as ppm free chlorine. mV and ppm free chlorine are the two

most commonly used sanitizer measurement units in water quality

management.

1. ORP Mode Selection

ORP mode selection allows the user to choose to display ORP readings

in either ORP millivolts (mV) or as converted to parts per million (ppm)

free chlorine. Use ORP to measure the effectiveness of all oxidizing

and reducing sanitizers, such as ozone, bromine and chlorine. Use Free

Chlorine in systems sanitized by chlorine only. To select between ORP

and Free Chlorine modes:
1.

Press

.

2.

Press and hold

for approximately 3 seconds.

The current preference for ORP units of measure is displayed. Factory

setting for this preference is mV.

3.

Press the

or

keys to toggle between mV

(standard ORP mode) and ppm (free chlorine measurement

mode). The setting chosen is displayed.

PPM

4.

Press any parameter key to exit ORP unit preference selection

or let the unit time out. ORP unit preference will be saved.

2. Measuring ORP

1.

Ensure the 9P is in ORP mode (ref. ORP Mode Selection, pg. 21).