Myron L 9PTK User Manual

Page 36

Advertising
background image

32

to the conductivity cell to release any CO

2

bubbles clinging

to the conductivity cell/cell cup extender wall. This effectively

reduces any bicarbonate alkalinity in the sample and prepares

it for titration. After 120 seconds, “add” alternating with “h1

1” will display. (The “1” indicates the first addition of sodium

hydroxide.) Then “agit” alternating with “hOld” will display.

13.

Change the pipette tip.

14.

Using the pipette, add 100µL of Reagent H1 to the cell,

close the cell extender cap securely to prevent any solution

from spilling out, then agitate and hold. Repeat as prompted.

P

r

es cal” will display.

15.

Press cal. “add” alternating with “h1 2” will display. (The

2” indicates the second addition of sodium hydroxide.) Then

agit” alternating with “hOld” will display.

16.

Tap the cell extender cap to dispel any solution clinging to the

cap back into the cell. Carefully open the cell extender cap so

as not to spill any solution.

17.

Using the pipette, add a second 100 µL of Reagent H1 to the

cell, close the cell extender cap securely to prevent any solution

from spilling out, then agitate and hold. Repeat as prompted.

P

r

es cal” will display.

18.

Press cal to advance to first titration prompt. “add” alternating

with either “edta lc” (Reagent H2) or “edta hc” (Reagent

H3) will display; then “agit t1” alternating with “hOld” will

display. The number after “t” indicates which titration point you

are on. The number “1” here indicates this is the first titration

point. Subsequent titration points are sequenced numerically: 2,

3, 4, etc.

19.

Tap the cell extender cap to dispel any solution clinging to the

cap back into the cell. Carefully open the cell extender cap so

as not to spill any solution.

20.

Change the pipette tip.

21.

Using the pipette, add 100µL of Reagent H2 if “edta lc” is

displayed, or H3 if “edta hc” is displayed.

22.

Close the cell extender cap securely to prevent any solution

from spilling out. Then agitate and hold. Repeat as prompted.

P

r

es cal” will display.

23.

Press cal to advance to the next titration point. Continue adding

100µL of Reagent H2 or H3 using the pipette, agitating and

holding the unit as prompted, then pressing cal as prompted.

Advertising