Myron L 9PTK User Manual

Page 20

Advertising
background image

16

unit as prompted, then pressing cal as prompted. An alkalinity

value will display when the titration measurement is complete.

14.

Note the value for your records or press ms to store the value.

You can begin another Alkalinity titration by pressing alK.
The unit will automatically power off after the period of inactivity

defined in the AUTO OFF setting (ref. AUTO OFF, pg. 43).

F. Measuring Hardness

1.

Hardness Unit Selection

The 9P offers the ability to set the hardness unit preference to either

PPm” CaCO

3

or grains of hardness. To change hardness units, press

harD, then press and hold cal down until hard and sel are

displayed. Press UP or DOwn to toggle between “PPm” and grains

(no units are displayed for grains). Press cal to accept. This also sets

the hardness unit preference for the LSI Titration function and the LSI

Calculator simultaneously.
2.

Hardness Titration Procedure

Prepare the materials required for titration: Reagent: C1, Reagent: H1,

Reagent: H2, Reagent: H3, cell extender, T-plunger, 100 µL pipette with

a clean tip installed, and two extra clean tips.
nOte: If you do not complete each titration point within the 3-minute

timeout period, the unit exits the Hardness function. Press harD to

start the titration process over from step 1. Between titrations, the unit

will automatically power off after the period of inactivity defined in the

AUTO OFF setting (ref. AUTO OFF, pg. 43).
1.

Ensure pH protective cap is secure to avoid contamination.

2.

Install the cell extender by pushing the base of the cell extender

into the conductivity cell until it is fully seated.

3.

Rinse the conductivity cell cup and cell extender 3 times with

solution to be measured.

4.

Refill cell cup and cell extender with sample solution.

5.

Insert T-plunger into open cell extender until the “T” rests flush

on the lip of the cell extender. This will cause fluid beyond what

is required for titration to overflow the cell extender.

6.

Rotate the T-plunger from side to side, allowing the arms of the

T-plunger to slide along the rim of the cell extender.