A. Conductivity Characteristics

When taking conductivity measurements, the Solution Selection

determines the characteristic assumed as the instrument reports what

a measured conductivity would be if it were at 25°C. The characteristic

is represented by the tempco, expressed in %/°C. If a solution of 100

µS at 25°C increases to 122 µS at 35°C, then a 22% increase has

occurred over this change of 10°C. The solution is then said to have a

tempco of 2.2 %/°C.

Tempco always varies among solutions because it is dependent on their

individual ionization activity, temperature and concentration. This is why

the Ultrameter III features mathematically generated models for known

salt characteristics that also vary with concentration and temperature.

B. Finding the Tempco of an Unknown Solution

One may need to measure compensated conductivity of some solution

unlike any of the 3 standard salts. In order to enter a custom fixed tempco

for a limited measurement range, enter a specific value through the User

function. The tempco can be determined by 2 different methods:

1.

Heat or cool a sample of the solution to 25°C, and measure its

conductivity. Heat or cool the solution to a typical temperature

where it is normally measured. After selecting User function,

set the tempco to 0 %/°C as in Disabling Temperature

Compensation, pg. 24 (No compensation). Measure the new

conductivity and the new temperature. Divide the % decrease

or increase by the 25°C value. Divide that difference by the

temperature difference.

2.

Heat or cool a sample of the solution to 25°C, and measure

its conductivity. Change the temperature to a typical measuring

temperature. Set the tempco to an expected value as in User

Programmable Temperature Compensation, pg. 24. See if the

compensated value is the same as the 25°C value. If not, raise

or lower the tempco and measure again until the 25°C value is

read.

C. Finding the TDS Ratio of an Unknown Solution

Once the effect of temperature is removed, the compensated

conductivity is a function of the concentration (TDS). There is a ratio

of TDS to compensated conductivity for any solution, which varies

with concentration. The ratio is set during calibration in User mode

as in User Programmable Conductivity to TDS Ratio, pg. 25. A truly

unknown solution has to have its TDS determined by evaporation

and weighing. Then the solution whose TDS is now known can be

measured for conductivity and the ratio calculated. Next time the same

solution is to be measured, the ratio is known.