3. User Programmable Conductivity to TDS Ratio

This feature allows you to select a custom conductivity to TDS conversion

ratio within the range of 0.20-7.99 for User mode measurements.

To determine the conversion ratio for a custom solution of known

TDS ppm value, measure the solution conductivity at 25ºc with the

Ultrameter III and divide the ppm value by the µS value. For example,

a solution of known 75 ppm TDS and measured 100 µS conductivity at

25ºC would have a conversion ratio of 75/100 or 0.75. Enter the new

conversion ratio as follows:

a.

While in User mode, press

.

b.

Press

twice (to skip over tempco adjustment) and

ratiO” will appear (see Figure 4).

c.

Adjust with

or

key until new conversion ratio is

displayed.

d.

Press

twice (to skip over calibration adjustment) to

accept new conversion ratio. You are now ready to measure

samples with the new conductivity/TDS ratio.