29

d.

Press

to accept the factory calibration setting.

5. Alkalinity Calibration

Prepare the materials required for titration: Alkalinity Standard Solution

100PPM, Reagent: A1, cell extender, T-plunger, and 100 µL pipette with

a clean tip installed.
nOte: If you do not complete each titration point within the 3-minute

timeout period, the unit exits the Alkalinity function. Between titrations,

the unit will automatically power off after the period of inactivity defined

in the AUTO OFF setting (ref. AUTO OFF, pg. 43). Press alK to start the

titration process over from step 1.
If you make a gross error in your titration method, e.g., you forget to add

sample to the cell, “err” will display. Start the titration over from Step 1

to obtain a reading.
1.

Ensure pH protective cap is secure to avoid contamination.

2.

Install the cell extender by pushing the base of the cell extender

into the conductivity cell until it is fully seated.

3.

Rinse the conductivity cell cup and cell extender 3 times with

Alkalinity Standard 100PPM.

4.

Refill cell cup and cell extender with Alkalinity Standard 100PPM.

5.

Insert T-plunger into open cell extender until the “T” rests flush

on the lip of the cell extender. This will cause fluid beyond what

is required for titration to overflow the cell extender.

6.

Rotate the T-plunger from side to side, allowing the arms of the

T-plunger to slide along the rim of the cell extender.

7.

Carefully remove the T-plunger so as not to flick or spill remaining

sample. Tap the T-plunger on the side of the cell extender to

remove sample that may be stuck to the T-plunger.