Myron L 9PTK User Manual

Page 21

Advertising
background image

17

7.

Carefully remove the T-plunger so as not to flick or spill remaining

sample. Tap the T-plunger on the side of the cell extender to

remove sample that may be stuck to the T-plunger.

8.

Press

. “ca” alternating with “tds” displays along with

PPm” sample value. “P

r

es cal” displays when the reading

stabilizes.

9.

Press cal. “add” alternating with “c1” will display; then “agit

alternating with “hOld” will display.

10.

Using the pipette, add 100µL of Reagent C1 to the sample,

close the cell extender cap securely, then agitate and hold.

Repeat as prompted. “OPe

n

caP” alternating with “P

r

es cal

will display.

11.

Tap the cell extender cap to dispel any solution clinging to the

cap back into the cell. Carefully open the cell extender cap so

as not to spill any solution.

12.

Press cal to start 120 second timer. The display will count

down from “120” to “0”, allowing time for CO

2

to escape. Using

your index finger, firmly tap the side of the instrument closest

to the conductivity cell to release any CO

2

bubbles clinging

to the conductivity cell/cell cup extender wall. This effectively

reduces any bicarbonate alkalinity in the sample and prepares

it for titration. After 120 seconds, “add” alternating with “h1

1” will display. (The “1” indicates the first addition of sodium

hydroxide.) Then “agit” alternating with “hOld” will display.

13.

Change the pipette tip.

14.

Using the pipette, add 100µL of Reagent H1 to the cell,

close the cell extender cap securely to prevent any solution

from spilling out, then agitate and hold. Repeat as prompted.

P

r

es cal” will display.

15.

Press cal. “add” alternating with “h1 2” will display. (The

2” indicates the second addition of sodium hydroxide.) Then

agit” alternating with “hOld” will display.

16.

Tap the cell extender cap to dispel any solution clinging to the

cap back into the cell. Carefully open the cell extender cap so

as not to spill any solution.

17.

Using the pipette, add a second 100 µL of Reagent H1 to the

Advertising