Copy tare, Using option “sample selection, Copy tare 25.6. using option “sample selection – RADWAG MYA 5.3Y.F1 Microbalance User Manual

Page 113

Advertising
background image

- 113 -

On previewing a specific weighing, a window is opened with data and values

on the weighing process. Numbers displayed next to weighing data <B>

denote measuring cycles.

Data on a specific series can be printed or exported to a file.

25.5. Copy tare

The option enables copying once selected tare value in a series to all samples

which do not have tare value assigned, and for which differential weighing

process (status other than Weighing B) is not completed.

Means of operation:

Press key

Which displays list of samples for which

tare value is assigned.

Press tare value that should apply to

other samples in a series (which do not

have tare applied).

The software automatically applies

selected tare value to data of these

samples.

25.6. Using option “SAMPLE SELECTION”

The option enables selecting a sample that is prepared for weighing, with no

need to weigh samples is a sequence as suggested by the software. The option is

available only after initiating a given process.

Means of operation:

While in differential weighing mode, on selecting a series, press procedure

selecting key, e.g. <Weighing A>.

Then press the following key:

which opens a list of samples for which

the selected procedure can be carried

out.

Select a sample to be weighed from

displayed list.

This option is particularly useful if a

series contains a large number of

samples, and the user should weigh a

sample that is at the end of a series.

Advertising