- 27 -

set to SAVE or SAVE & ALERTS mode, and there is no physical connection

between a balance and a THB module, or the parameters for communicating

the module with the balance are set incorrectly. In such case plug a THB

module to balance’s ports COM 1 or COM 2 and set appropriate cooperation

parameters, such as specified on module’s data plate.

10.7. Additional parameters on weighing process

Changes in settings of functions related to weighing process can in some

extend program balance reactions.

Procedure:

1. Press grey coloured workspace

2. Balance display indicates menu: Settings, Keys, Information, Printouts,

Profile

3. Press one of available submenu options and select item for modification,


Menu <Settings> - additional options related to weighing mode

Menu <Keys> - defining quick access keys
Menu <Information> - supplementary data on weighing process displayed in

the grey coloured workspace

Menu <Printouts> - selection of printout type
Menu <Profile> - selection of profile, active during balance operation

Menu <SETTINGS> - contains supplementary data on weighing process, such

as:

Tare mode

SINGLE,

mass value stored in balance memory on single pressing of TARE key, the

following pressing of TARE key determines new value of tare. Selection of

product or packaging with assigned tare value causes automatic deleting

of previously assigned tare value;

SUM OF ACTIVE,

totalizing applied tare values for a product or a packaging (resulting

from selecting a product or a packaging from corresponding database)

with a possibility of increasing determined tare value by manual tare

introduced manually using balance’s numeric keyboard. On following

determining of tare value for a product or a packaging, the tare value

introduced manually is deactivated.

SUM OF ALL,

totalizing all introduced tare values.

AUTOTARE

Means of operation:

Each first stable measurement result is tarred. The NET pictogram

flashes on the display. The operator can determine net mass of weighed

load. On taking off the load from balance’s weighing pan, and as the

Advertising