22.4. Using formulas in weighing

In activating FORMULATION mode, depending on settings, user can move to

formula making process either on selecting a formulation from corresponding

database or weigh a mixture “manually”.

Preparing mixtures is carried out by:

Preparing a formulation that is not saved in the Database of Formulas –

i.e. “manually”,

Preparing a formulation that is saved in the Database of Formulas,

Preparing multiple repetitions of a formulation that is saved in the

Database of Formulas – multiplier function,

Preparing a formulation that is saved in the Database of Formulas, with

determining target mass of complete formulation.

Caution:

Using the above options requires enabling corresponding quick access keys in the

bottom bar and setting corresponding options in the formulation mode.

On selecting a formulation, weigh each of its ingredients. On stabilization of

each measurement result press <

> key. Mass of each ingredient after

pressing <

> is zeroed, i.e. it is accepted for the total mass of the

formulation.

Pressing <

> key cancels the formula making process. If pressed, enables

selecting another formulation to making process.

PROCEDURE 1 – Preparing a formulation that is not saved in the Database

of Formulas – i.e. “manually”.

Follow hints and commands displayed by the balance’s software

Press <

Formulation without a name> key in the quick access key

bottom bar,

Give name to the new formulation,

The software automatically skips to selecting ingredients of a formulation,

Select an ingredient (new or from database of products),

The software returns to the main window,

Weigh set quantity of an ingredient,

Accept process by pressing <

> key,

The software moves to selecting the following ingredient,

Proceed as previously to weigh all required ingredients of a formulation,

After weighing the last ingredient, select one of below options:

<Save and close> - carried out formula making process is saved in the

database of formulation reports and automatically completed,

<Close> - carried out formula making process is completed without

saving a final report in the database of formulation reports,

On selecting one of above options, the software finishes formula making

process and returns to displaying the main window of Formulation mode.