10. The balance automatically calculates the density of air, and indicates its

value on the display,

11. Press <

> key to finish the procedure.

The value of determined air density is automatically assigned to field <Air

density> in menu <Settings/ Air buoyancy compensation> in working mode:

weighing.

20.4. Additional setings of density mode

The additional settings enable adjusting the working mode to user’s needs and

requirements. To access the setting follow below procedure:

Procedure:

1. Press grey coloured workspace,

2. The display indicates menu of: Settings, Keys, Information, Printout

and Profile,

3. Press <Settings> key,

4. The display indicates functions related to density determining

process, such as:

o

Request sample number

Available settings:

NO – data on sample number is not required; applicable in case of

a single measurement

YES – before each measurement a window requesting inserting

sample number is automatically displayed; applicable in case of a

series of measurements

o

Measuring unit: the balance enables choosing between the

following units: [g/cm

3

], [kg/m

3

], [g/l]. Selected unit is activated

for all density determining options and also in final printout.

o

Assign density to a product: YES/NO – if option <YES> is

selected, the software automatically assigns the value of

determined density of solid or liquid to field <Density> in a

product record of the database of products. To use this function,

before starting the density determining process, using database of

product select a product for which the density procedure is carried

Advertising