Acting according to the command, place an ordered weight/standard on

balance’s weighing pan and press

key,

On completing the adjustment procedure the balance indicates a message

box:

On confirming the message box with

key, the balance returns to

weighing mode.

11.3. User adjustment

User adjustment is carried out with an optional standard, which mass ranges

between 0,15 Max and Max. Adjustment procedure is compatible with the

external adjustment process, but before its start the software opens a

message box for entering mass of a standard used for user adjustment.

Caution:

User adjustment is available only in balances which are not subject to conformity

assessment (verification).

In order to start user adjustment, enter submenu < Adjustment >, and select
option: “ User adjustment”. Then follow the commands indicated on

balance’s display.

11.4. Adjustment test

<Adjustment test> function enables comparing the result of internal

automatic adjustment with the value of internal weight saved in balance’s

factory parameters. The comparison is used for determining balance’s

sensitivity drift over time.

11.5. Automatic adjustment

This menu option is used for selecting a factor which determines start of

automatic adjustment process. Accessible options:

None – automatic adjustment disabled