26.2. Additional settings of the SQC mode

The additional settings enable adjusting the working mode to user’s needs and

requirements. Access to these settings is described below.

Procedure:

1. Press grey coloured workspace

2. The display indicates menu of: Settings, Keys, Information, Printout ,

Batch and Profile,

3. Press <Settings> key,

4. The display opens a window with set of available options

The SQC mode is related to the following settings:

Request batch number

Setting parameter value to <YES> causes that the software requests entering

a batch number of a product before starting each control process.

Procedure:
Select a parameter: “

Batch number” which opens an editing window with

available settings,

Select required batch quantity, after which the software returns to displaying

the previous window.

Declaring batch quantity

Quantity (size) of a batch is a number of samples intended to be measured.

Procedure:
Select a parameter: “

Batch quantity” which opens an editing window

<Batch quantity> with a numeric keyboard,
Enter required batch quantity and accept by pressing

key.

For the other settings follow the principle given in point 10.8 of this user

manual “Additional parameters on weighing process”.

26.3. Carrying out a control process

Before starting a control process select a product to be controlled from the
database of products using a quick access key <

Product>. After product

selection the control is ready to be initiated.
In order to start a control press

key located in the bottom bar of the display.

The software automatically opens an editing window with an on-screen keyboard

for entering batch number of a product to be controlled. After entering and

accepting the batch number, the software goes to another step of the process.

As the control process is in progress, the software carries out a real time

analysis of the results, inserts them in the corresponding fields of the display and

simultaneously informs a user in the results of carried out control.