Ambient conditions, Packages, Ambient conditions 27.14. packages – RADWAG MYA 5.3Y.F1 Microbalance User Manual

Page 135

Advertising
background image

- 135 -

measuring range of a balance (the <TARE>

button applied), as well as on weighing

samples without the packaging (the

<TARE> button is not applied).

Example no. 3 a balance AS 220.3Y:

No. Tare value

Minimum

sample

Means of operation

1

0.0000 g

0.2500 g The minimum sample refers to all net

weighments that are weighed without a

packaging (the <TARE> button is not

applied).

The software identifies this setting as the

minimum sample is valid only for weighing

samples without a packaging. If tarring

option is applied then the software disables

the pictogram informing on minimum

sample in use, and renders the weighment

as carried out with tare applied, that does

not have a minimum sample criterion

determined.

The user can preview entered data, but data editing is disabled.

27.13. Ambient conditions

Database of ambient conditions contains parameters related to measurement

of ambient conditions at a workstation. Depending on balance’s configuration,

a report may contain data on temperature, humidity and value of atmospheric

pressure. In case a THB ambient conditions module is connected to a balance,

records from the module are also saved.

Procedure

Enter submenu < Databases>, and press <

Ambient conditions >

field

Press field of a desired report record, unless it is visible on the top of the

list, wind the list forward using navigating keys

Report’s name consists of date and time

Caution:

It is possible to use an option for searching a report.

27.14. Packages

Database of packages contains list of packages for which a name, code and

mass value are specified. During weighing process, if a name of packaging is

selected from the database, the tare value is automatically applied. The

display indicates mass value with minus sign.

Advertising