Format of a response to computer command – RADWAG MYA 5.3Y.F1 Microbalance User Manual

Page 167

Advertising
background image

- 167 -

SS

Result release

IC

Internal calibration

IC1

Lock automatic internal adjustment

IC0

Unlock automatic internal adjustment

NB

Give balance factory number

K1

Lock balance keyboard

K0

Unlock balance keyboard

BP

Start BEEP signal

OMI

List available working modes

OMS

Set working mode

OMG

Give current working mode

PC

Send all implemented commands

Caution:

1. Each command must end with CR LF characters;

35.2. Format of a response to computer command

On receipt of a command, the terminal sends a response in one of the

following formats:

XX_A CR LF

Command understood and in progress

XX_D CR LF

Command carried out (appears only after the

command XX_A)

XX_I CR LF

command understood but not accessible at this

moment

XX _ ^ CR LF

Command understood but max range is exceeded

XX _ v CR LF

Command understood but min range is exceeded

XX _ OK CR LF

Command carried out (completed)

ES_CR LF

Command not recognized

XX _ E CR LF

An error occurred on command carrying out (time

limit exceeded while waiting for stable measurement

result (time limit is balance’s characteristic

parameter)

XX - stands for a name of sent command

_ - substitutes spaces

COMMANDS DESCRIPTION

Zero the balance

Format: Z CR LF