Checkweighing – RADWAG MYA 5.3Y.F1 Microbalance User Manual

Page 60

Advertising
background image

- 60 -

17. CHECKWEIGHING

Working mode<

Checkweighing > enables controlling sample mass

using two threshold (Low and High). Usually it is accepted, that mass

indication is correct if it is included within the thresholds’ values.

Working mode activating procedure

While in the main window, press

soft key located in the upper bar of the

display, which opens a submenu <Working modes> with selection of

available working modes,

Select <

Checkweighing> mode, the software automatically returns to

the main window and displays

pictogram in the upper bar.

The grey coloured workspace contains the following data:

o

Supplementary unit

o

Low threshold

o

High threshold

o

Difference – denoting the “distance” of current measurement result from

the center of Lo-Hi section.

On selecting the checkweighing mode the display contains the

following quick access keys in the bottom bar:

1. Setup – access to balance’s menu,

2. Print header – print of data declared in the header,

3. Print footer – print of data declared in the footer,

4. Database of products – selection of products from corresponding

database,

5. Set tare – field for setting numeric value of tare,

6. Checkweighing thresholds – for declaring the values of Low and High

threshold.