Adjustment history – RADWAG MYA 5.3Y.F1 Microbalance User Manual

Page 37

Advertising
background image

- 37 -

dashes,

signature,

non-standard printout.

11.9. Adjustment history

Contains data on all carried out adjustment processes. The record is carried

out automatically. Each entry on adjustment comprises basic data on

completed process. Balance menu enables displaying the list of completed

adjustment processes, and each report is printable.

Printing a report on adjustment process. Enter submenu < Adjustment >,

and: <Adjustment history>, then select for printing adjustment entry from a
list. On displaying details of a record, press print soft key < > on display’s

upper bar.

Hint:

If memory for records on completed adjustments is full, then the oldest record is

automatically erased.

If internal procedures of an organization require maintaining complete

documentation from all carried out adjustment processes, then the list with

records on adjustment should be periodically printed and filed.

Searching for adjustment record

Balance enables searching for a specific record from completed adjustment

processes:
– Press search icon

and insert date of adjustment process.

Exporting data on completed adjustment process

Connect a data storage device to balance’s USB port. Press<Data export> key

located in the upper right corner of balance’s display. The process is fully

automatic, and on its completion, a file with extension *.tdb is saved on a

data storage device connected to the USB port. The file is editable using Excel

spreadsheet or a text editor.

Advertising