BNC SAM 935 Portable Gamma Spectroscopy System User Manual

Page 50

Advertising
background image

46

SAM 935™ Instruction Manual

2.

If you know what the value should be, press F3 to manually edit the value:

When finished, press E

NTER

to save the new value.

-

OR -

3.

To perform the dose rate calibration, press F2. The SAM 935 will tell you that you need a Cs137
source of known activity and date to do this:

4.

Press E

NTER

.

5.

Enter the activity of the calibration standard at the
next screen and press E

NTER

:

6.

Enter the standard’s calibration date at the next
screen and press E

NTER

:

7.

You will be prompted to place the Cs137 standard 20
centimeters (7.87 in.) away from the SAM 935’s detector. Press OK (F2) when you have done so.

You will see a spectrum resolve on screen as the calibration is taken.

Contact your supplier if needed to get the value to
be entered manually here.