3.2 Mounting base tray and bottle holder (optional)

■■■■■■■■■■■■■■■■■■■■■■

10

■■■■■■■■

2.896.0030 896 Professional Detector – Conductivity & Amperometry

If a 896 Professional Detector – Conductivity & Amperometry is to be set
up underneath an 881 Compact IC pro or an 882 Compact IC plus, then
the System Connector (6.2061.120) must be used instead of the base tray
of the IC instrument; this adjusts the larger surface area of the 896 Profes-
sional Detector – Conductivity & Amperometry to the smaller surface area
of the 881 or 882, respectively.

The 896 Professional Detector – Conductivity & Amperometry cannot be
set up above an 881 Compact IC pro or an 882 Compact IC plus.

3.2

Mounting base tray and bottle holder (optional)

The 6.2061.110 base tray and 6.2061.100 bottle holder protect IC instru-
ments from dust, dirt and leaking fluids. If more than one instrument of
the Professional IC family is used, these can be set up in one or more
stacks. We recommend that a base tray and a bottle holder be mounted
for each stack of IC instruments.

Bottle holder and base tray must be removed or mounted every time one
of the following instruments is to be mounted on or under an 850 Profes-
sional IC instrument:

One or more 872 Extension Module.

An 887 Professional UV/VIS Detector.

An 896 Professional Detector.

or another instrument of the same support surface.

3.2.1

Removing / mounting the base tray

The base tray must be removed, in case you want to place another instru-
ment under the IC instrument. Proceed as follows:

Removing the base tray

Before you can remove the base tray, the following preconditions must be
met:

The instrument is switched off.

The bottle holder is cleared.

All connections at the rear of the instrument are disconnected.

There are no loose parts in the instrument.

To remove the base tray, you need a 3 mm hexagon key (6.2621.100).