6.9 Systematic error diagnostics

■■■■■■■■■■■■■■■■■■■■■■

42

■■■■■■■■

2.896.0030 896 Professional Detector – Conductivity & Amperometry

6.9

Systematic error diagnostics

If the causes of a malfunction cannot be found among the problem
descriptions in the above chapters, then proceed systematically as follows:

Systematic error diagnostics

1 Restarting the instrument and the software

Switch off the instrument.

Close and restart MagIC Net™.

Switch the instrument back on again.

If the problem has not yet been localized, continue with Step 2.

2 Performing instrument test with dummy cell

(see Chapter 4.1, page 20)

If the problem has not yet been localized, continue with Step 3.

3 Checking software settings

Check the method parameters of the detector and reset them to
values that you know will function.

Check the measuring range and reset it to values that you know
will function or select a larger measuring range.

Check manual changes to the settings and reset them to values
that you know will function.

Check manual settings in the time program and reset them to val-
ues that you know will function.

If the problem has not yet been localized, continue with Step 4.

4 Cleaning the measuring cell

Switch off the measuring cell.

Remove the measuring cell.

Clean the measuring cell (see measuring cell manual).

Polish the working electrode (see the leaflet for the working elec-
trode).

Reinsert the measuring cell.

If the problem has not yet been localized, continue with Step 5.

5 Replacing the reference electrode

If the problem has not yet been localized, continue with Step 6.