3.3 Conductivity detector

■■■■■■■■■■■■■■■■■■■■■■

16

■■■■■■■■

2.896.0030 896 Professional Detector – Conductivity & Amperometry

Note

In order to prevent unnecessary peak widening after separation, the
connection between the outlet of the separation column and the inlet
in the detector should be kept as short as possible.

Connecting the detector input capillary to the separation
column

1 Connect the detector input

Fasten the detector input capillary (4-1) using a short Peek pres-
sure screw (6.2744.070) (4-2)
directly to the output of the col-
umn (4-3).

1

2

3

Figure 4

Connection detector – separation column

1

Detector input capillary

2

PEEK pressure screw, short
(6.2744.070)

3

Separation column

Connecting the detector input capillary to the suppressor

1 Connect the detector input

Connect the detector input capillary (5-1) and the capillary of the
suppressor (5-2) labeled with out to one another using one cou-
pling (6.2744.040) (5-3) and two short PEEK pressure screws
(6.2744.070) (5-4).