Figure 5, Connection detector – suppressor, Figure 6 – Metrohm 896 Professional Detector – Conductivity & Amperometry User Manual

Page 25: Connection detector – mcs, 2) labeled with

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

2.896.0030 896 Professional Detector – Conductivity & Amperometry

■■■■■■■■

17

1

2

3

4

Figure 5

Connection detector – suppressor

1

Detector input capillary

2

Suppressor output capillary
labeled with out.

3

Coupling (6.2744.040)

4

PEEK pressure screws, short
(6.2744.070)

Connecting the detector input capillary to the MCS

1 Connect the detector input

Fasten the detector input capillary (6-1) with one long PEEK pres-
sure screw (6.2744.090) (6-2) to the output of the MCS (6-3).

3

2

1

Figure 6

Connection detector – MCS

1

Detector input capillary

2

PEEK pressure screw, long
(6.2744.090)

3

MCS output