4.7 Connecting the electrode connection cables

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

2.896.0030 896 Professional Detector – Conductivity & Amperometry

2 Remove the purge stopper

Remove the stopper from the Purge connector.

3 Deaerate the measuring cell

Observe the eluent that emerges through the deaeration opening.
Wipe up liquid with a cloth.

Once no more air bubbles are visible, screw the stopper back on the
Purge connector and tighten it by hand.

4 Switch off the high pressure pump in MagIC Net™.

4.7

Connecting the electrode connection cables

Connecting the electrode connection cables to the detector

Precondition: The electrode connection cables are connected to the meas-
uring cell (see chapter installation in the manual of the measuring cell).

1 Plug in the straight plug of the working electrode connection cable

(red sleeve) into the WE socket of the detector.

2 Plug in the straight plug of the reference electrode connection cable

(black sleeve) into the RE socket of the detector.

3 Plug in the straight plug of the auxiliary electrode connection cable

(blue sleeve) into the AE socket of the detector.