4 power connection, 5 leak sensor, 6 ambient conditions – Metrohm 896 Professional Detector – Conductivity & Amperometry User Manual

Page 54: Power connection, Leak sensor, Ambient conditions

Advertising
background image

7.4 Power connection

■■■■■■■■■■■■■■■■■■■■■■

46

■■■■■■■■

2.896.0030 896 Professional Detector – Conductivity & Amperometry

Remote

Via Remote Box

Analog output

With 891 Professional Analog Out

Output voltage

0…1000 mV

Full scale

Can be adjusted within the digital signal range

Offset

Can be adjusted within the digital signal range

System standby

Automatic function test at start-up

Leak sensor

Temperature stability monitoring

Output channels

Current rating

Charge

GLP conformity

Yes, optional

7.4

Power connection

Required voltage

100 - 240 V ± 10% (autosensing)

Required fre-
quency

50 - 60 Hz ± 3 Hz (autosensing)

Power consump-
tion

65 W for typical analysis application

25 W standby (conductivity detector to 40 °C)

Power supply unit

Up to 300 W maximum, electronically monitored

internal fuse 3.15 A

7.5

Leak sensor

Type

Electronic, no calibration necessary

7.6

Ambient conditions

Operation

Ambient tem-
perature

+5…+45 °C

Humidity

20…80 % relative humidity

Storage

Ambient tem-
perature

–20…+70 °C

Transport

Ambient tem-
perature

–40…+70 °C