Metrohm 896 Professional Detector – Conductivity & Amperometry User Manual

Page 29

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Start-up

2.896.0030 896 Professional Detector – Conductivity & Amperometry

■■■■■■■■

21

Plug in the straight plug of the working electrode connection
cable (red sleeve) into the WE socket of the detector.

Plug in the straight plug of the reference electrode connection
cable (black sleeve) into the RE socket of the detector.

Plug in the straight plug of the auxiliary electrode connection
cable (blue sleeve) into the AE socket of the detector.

3 Insert the dummy cell

Place the dummy cell into the tray of the detector.

Attach the front lid.

Note

The metal parts of the cable plugs must not touch the front lid.

4 Adjust settings in MagIC Net™

In the program part Method, create a new method for the instru-
ment test with the dummy cell.

Select the detector and add it as a new device.

Select the DC mode.

Set the following parameters for the DC mode:

DC potential: 0.8 V
Range: Auto
Damping: off

Add an analysis for the detector channel Current.

Save the method.

In the program part Workplace:

Load the method.

In the Watch window, display the Current channel with at least
3 decimal places.

5 Carry out the test

In the program part Manual:

Switch on the cell
After no more than 1 minute, the detector signal should level off
at 2.667 nA ± 7 %. Noise should not exceed 0.005 nA.