Metrohm 896 Professional Detector – Conductivity & Amperometry User Manual

Page 34

Advertising
background image

4.5 Testing the measuring cell

■■■■■■■■■■■■■■■■■■■■■■

26

■■■■■■■■

2.896.0030 896 Professional Detector – Conductivity & Amperometry

Note

Do not move the measuring cell for at least 5 seconds after having
inserted it.

During this time, data is read from the chip of the measuring cell
and written into the database. This process must not be interrup-
ted, because otherwise the data may be transferred incorrectly or
incompletely.

3 Test at low flow

In the program part Manual of MagIC Net™, set the flow of the
high pressure pump to 0.2 mL/min and start the high pressure
pump.

Watch the detector output capillary: Liquid must emerge from the
end of the detector output capillary.
If no liquid emerges from the end of the detector output capillary:

– Undo the capillary from the Out connection on the meas-

uring cell and check whether the end has been squeezed by
the pressure screw.

– Shorten the capillary and fasten once again to the Out con-

nector on the measuring cell.