5 testing the measuring cell, Testing the measuring cell – Metrohm 896 Professional Detector – Conductivity & Amperometry User Manual

Page 33

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Start-up

2.896.0030 896 Professional Detector – Conductivity & Amperometry

■■■■■■■■

25

4.5

Testing the measuring cell

To test the measuring cell, proceed as follows:

Testing the measuring cell

Preconditions:

The measuring cell is completely assembled (see measuring cell man-
ual).

The working electrode and the reference electrode are inserted (see
measuring cell manual).

1 Connect the measuring cell

Connecting the measuring cell input:

When the preheating capillary is used: Use a pressure

screw (6.2744.014) to fasten a piece of the PEEK capillary
(6.1831.010) to the Eluent to cell connector on the
detector.
Use a pressure screw (6.2744.014) to fasten the other end
to the In connector on the measuring cell.

If the preheating capillary is not used: Use a pressure screw

(6.2744.014) to fasten the detector input capillary directly
to the In connector on the measuring cell.

Connecting the measuring cell output:
Use a pressure screw (6.2744.014) to fasten the tested detector
output capillary to the Out connector on the measuring cell (see
"Testing the detector output capillary", page 24)
.

2 Insert the measuring cell

Insert the chip of the measuring cell into the cell holder so that you
can hear it locking in.