4 Start-up

2.896.0030 896 Professional Detector – Conductivity & Amperometry

■■■■■■■■

27

Observe the measuring cell: No liquid should emerge from the
body of the measuring cell.
If the measuring cell is leaking:

– Remove the measuring cell from the measuring cell holder.
– Remove all capillaries and cables.
– Check the seating of the pressure screw of the working

electrode and retighten it.

– Restore the capillary connections.
– Reinsert the electrode cables.
– Reinsert the measuring cell.
– Repeat the test.

4 Test at normal flow

In the program part Manual of MagIC Net™, raise the flow of
the high pressure pump to 1.0 mL/min.

Observe the measuring cell: No liquid should emerge from the
body of the measuring cell.

4.6

Deaerating the measuring cell

The cell must be deaerated in order to ensure that it contains no air bub-
bles.

The measuring cell must be deaerated after the installation and after each
subsequent opening of the cell.

Proceed as follows:

Deaerating the measuring cell

Prerequisites:

The high pressure pump is switched on and pumps the eluent through
the IC system to the measuring cell.

The measuring cell is switched off.

1 Deaerate the reference electrode chamber

Unscrew the nut on the RE connector and remove it.

Lift out the reference electrode.

Wait until the reference electrode chamber has filled with eluent.

Reinsert the reference electrode. Wipe up any emerging eluent
with a cloth.

Screw the nut on the reference electrode connector back on
tightly.