4 problems with sensitivity, General remarks regarding sensitivity fluctuations, Problems with sensitivity – Metrohm 896 Professional Detector – Conductivity & Amperometry User Manual

Page 46

Advertising
background image

6.3 General remarks regarding sensitivity fluctuations

■■■■■■■■■■■■■■■■■■■■■■

38

■■■■■■■■

2.896.0030 896 Professional Detector – Conductivity & Amperometry

6.3

General remarks regarding sensitivity fluctuations

Sensitivity fluctuations of up to 20% per week are normal for an
unchanged system in constant operation.

The sensitivity can increase to approximately twice as much for a short
time when new working electrodes are inserted or when the conditions
change.

6.4

Problems with sensitivity

Problem

Cause

Remedy

Declining sensitivity.

Measuring cell – Auxiliary
electrode contaminated.

Clean the auxiliary electrode (see measuring
cell manual).

Incorrect eluent in the ref-
erence chamber.

Remove purge stopper on the measuring cell,
wait until approx. 1 mL of eluent has emerged,
screw the purge stopper back in tightly.

Sample concentration is no
longer correct.

Replace sample and/or standard solution.

Temperature fluctuations.

Amperometric detector – Use preheating
capillary.

IC instruments – Use column oven.

Replace the measuring cell.

Use a measuring cell of the same type.

Use the same spacer.

Use the same electrodes.

Software – Measurement
potential incorrect.

Optimize the measurement potential.

Measuring cell – Working
electrode contaminated.

Clean the working electrode (see the leaflet for
the working electrode).

IC system – Eluent conta-
minated.

Apply new eluent.

IC system – pH of the elu-
ent has changed.

Check the pH of the eluent and optimize it if
necessary.

Advertising