6.7 Problems with the chromatogram

■■■■■■■■■■■■■■■■■■■■■■

40

■■■■■■■■

2.896.0030 896 Professional Detector – Conductivity & Amperometry

Problem

Cause

Remedy

No measuring sig-
nal.

IC system – No mains cur-
rent.

Check mains connection and voltage.

Peaks cut off at the
top.

Measuring range too
small.

Set a less sensitive measuring range.

Reduce peak height, e.g. by means of sam-
ple dilution.

6.7

Problems with the chromatogram

Problem

Cause

Remedy

Peak drift with
sugar analysis.

Carbonate absorption in
the eluent.

Use trap-column Metrosep CO3 Trap 1
(6.1015.300).

Peaks have poor res-
olution.

IC system – Diminished
separating efficiency of the
separation column.

Regenerate the separation column (see
Chapter Handling and maintenance in the
manual for the IC instrument).

Replace the separation column (see Chap-
ter Handling and maintenance
in the man-
ual for the IC instrument).

IC system – Eluent is old.

Apply new eluent.

Ionic strength of the sam-
ple or pH of the sample
deviates greatly from the
eluent.

Dilute the sample or optimize the pH of the
sample.

Absorption of analyte at
the electrodes.

Use a suitable combination of electrodes and
eluent.

Unexpected change
to the retention
times in the chroma-
tograms.

IC system – Diminished
separating efficiency of the
separation column.

Regenerate the separation column (see
Chapter Handling and maintenance in the
manual for the IC instrument).

Replace the separation column (see Chap-
ter Handling and maintenance
in the man-
ual for the IC instrument).

IC system – Eluent is old.

Apply new eluent.

Ionic strength of the sam-
ple or pH of the sample
deviates greatly from the
eluent.

Dilute the sample or optimize the pH of the
sample.