5 Operation and maintenance

2.896.0030 896 Professional Detector – Conductivity & Amperometry

■■■■■■■■

33

3 In MagIC Net™, set the maximum pressure of the high pressure

pump to 5 MPa.

4 Rinse the detector thoroughly with eluent.

5.3

Amperometric detector

5.3.1

Maintenance

Warning

When rinsing the detector without column, the pressure must not
exceed 5 MPa.

In order to ensure this, set the maximum pressure of the high pressure
pump to 5 MPa in MagIC Net™.

5.3.2

Preheating capillary maintenance

The preheating capillary can become blocked, e.g. if the IC system has
inadvertently been run dry.

To dissolve this blockage, proceed as follows:

Rinsing the preheating capillary

1 Remove the separation column

Remove the separation column from the IC system and replace with
a coupling (6.2744.040).

2 Adjust settings in MagIC Net™

In MagIC Net™, adjust the following settings:

Maximum pressure of the high pressure pump: 5 MPa

Flow rate: < 0.1 mL/min

3 Rinse the system with the same eluent as before the blockage or

with ultrapure water.

The eluent requires sufficient time to trickle through and dissolve the
crystals.

4 Do not increase the flow rate until the pressure has stabilized.