6 Troubleshooting

2.896.0030 896 Professional Detector – Conductivity & Amperometry

■■■■■■■■

39

6.5

Problems with the pressure

Problem

Cause

Remedy

Marked drop in
pressure.

IC system – Leak in the sys-
tem.

Check capillary connections and seal them.

Marked rise in pres-
sure.

IC system – Inline filter
blocked.

Replace filter platelets (see Chapter Handling
and maintenance
in the manual for the IC
instrument).

IC system – Separation col-
umn contaminated.

Regenerate the separation column (see
Chapter Handling and maintenance in the
manual for the IC instrument).

Replace the separation column (see Chap-
ter Handling and maintenance
in the man-
ual for the IC instrument).

Note: Samples should always be micro-filtered
(see Chapter Handling and maintenance –
Inline sample preparation
in the manual for
the IC instrument).

Amperometric detector –
Preheating capillary
blocked.

Preheating capillary maintenance (see Chapter
5.3.2, page 33)
.

Amperometric detector –
Detector output capillary
not free of blockage.

Test the detector output capillary (see Chapter
4.4, page 23)
.

6.6

Problems with the measuring signal

Problem

Cause

Remedy

Measuring signal
"overload".

Air bubble in the measur-
ing cell.

Deaerate the measuring cell (see Chapter 4.6,
page 27)
.

Measuring cell – Working
electrode damaged.

Replace working electrode.

Measuring cell – Measur-
ing cell not correctly con-
nected.

Check cable connections (see Chapter 4.7,
page 28)
.

Software – Measurement
potential incorrect.

Optimize the measurement potential.