7.1 Reference conditions

■■■■■■■■■■■■■■■■■■■■■■

44

■■■■■■■■

2.896.0030 896 Professional Detector – Conductivity & Amperometry

7 Technical specifications

7.1

Reference conditions

The technical data listed in this Chapter refers to the following reference
conditions:

Ambient tempera-
ture

+25 °C (± 3 °C)

Instrument status

> 40 minutes in operation (equilibrated)

7.2

Conductivity detector

Type

Microprocessor-controlled Digital Signal Processing (DSP technol-
ogy)

Intelligent detector with 6 reference chromatograms

Measuring range

0…15000 µS/cm without range switching

Noise

< 0.1 nS at 1 µS/cm

Deviations from
the linearity

< 0.1 % for conductivity values higher than 16 µS/cm

< 1 % for conductivity values lower than 16 µS/cm

Drift

< 0.2 nS/cm per hour

Measuring rate

10 measurements per second for optimum results without filtering

Resolution

0.0047 nS/cm

Baseline

Noise < 0.2 nS/cm typical for sequential suppression

Conductivity
detector

Cell volume

0.8 µL

Cell constant

Individual calibration data saved in the detector

adjustable in the range: 13.0…21.0 /cm

Electrodes

Ring-shaped electrodes made from stainless steel

Materials in
contact with
eluent

Chemically inert PCTFE

Maximum oper-
ating pressure

5.0 MPa (50 bar)

Cell tempera-
ture

20…50 °C in steps of 5 °C