4 testing the detector output capillary, Testing the detector output capillary – Metrohm 896 Professional Detector – Conductivity & Amperometry User Manual

Page 31

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Start-up

2.896.0030 896 Professional Detector – Conductivity & Amperometry

■■■■■■■■

23

To check whether the preheating capillary is free of both leaks and block-
ages, proceed as follows:

Testing the preheating capillary

1 Connect the detector input capillary

Use a pressure screw (6.2744.014) to fasten the detector input capil-
lary to the connector Eluent in on the detector.

2 Adjust settings in MagIC Net™

In the program part Manual of MagIC Net™, set the maximum
pressure of the high pressure pump to 5 MPa.

Set the flow to 0.1 mL/min.

Start the high pressure pump.

3 Observe the Eluent to cell connector

After a while, liquid must emerge from the Eluent to cell connector
(wipe up fluid with paper towel).

If no liquid emerges at the Eluent to cell connector, then the pre-
heating capillary is likely to be blocked. To eliminate the problem, see
Chapter Preheating capillary maintenance, page 33
.

4 Observe the pump pressure

Observe the pump pressure display in the program part Manual of
MagIC Net™.

A constant pressure should establish itself after a while.

4.4

Testing the detector output capillary

The detector output capillary must be of a certain length in order to be
able to generate sufficient backpressure. The required length is dependent
on the flow that has been set. The table 1 shows the recommended
lengths, as determined by the set flow.

Table 1

Recommended lengths for the detector output capillary

Flow

Capillary length (

0.25 mm)

2.0 mL/min

0.5…1.5 m

0.5…1.0 mL/min

1.0…2.5 m

0.25 mL/min

3 m