4.3 Faults and malfunctions

844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051

89

4.3 Faults

and

malfunctions

4.3.1 Error

messages

If any type of malfunction occurs during operation of the 844 UV/VIS

Compact IC, this is shown by error messages in the PC program, which

appear either in an error window or in the

SYSTEM STATE

window.

Follow the instructions listed in the Error window and close this

window with

<OK>

.

You will find further details of the error messages of the

SYSTEM STATE

window, their possible causes and the procedure for rectifying them in

the Software Instructions for Use «IC Net 2.3» Section 4.5.

4.3.2

Malfunctions and their rectification

If difficulties appear with The 844 UV/VIS Compact IC during analyses,

their causes are best investigated in the order separating column

high-pressure pump eluent connections. Several of The

malfunctions which may appear are listed in the following table with

details of possible causes and countermeasures.

Malfunction

Cause

Rectification

Baseline with high
noise level,
pulsation

• Contaminated pump

valves

• Defective piston seals

• Clean the valves (see

Section 4.2.4)

• Replace the piston seals

(see Section 4.2.4)

Drift of the
baseline

• Thermal equilibrium not

yet reached

• Leak in system

• Evaporation of organic

solvent in eluent

• Condition system with

heating switched on

• Check connections and

make leakproof

• Ensure better closure of

eluent supply vessel

Considerable
pressure drop

• Leak in system

• Check connections and

make leakproof

Considerable
pressure rise

• Contamination of the

filter in the 6.2821.120
Filter unit PEEK

• Blockage of the preco-

lumn

• Change of column

packing by injection of
contaminated samples

• Replace the 6.2821.130

Filter (see Section 2.3.5)

• Change precolumn (see

Section 4.2.3)

• Regenerate column (see

Section 4.1.1) or replace
column (see Section 4.2.3)

Note:
Samples should always
be microfiltered.

Chromatograms
with poor
resolution, change
in the retention
times

• Deterioration in

separation efficiency of
the IC column

• Regenerate column (see

Section 4.1.1) or replace
column