Hardware settings – Metrohm 844 UV/VIS Compact IC User Manual

Page 82

Advertising
background image

Table of contents

Peristaltic pump

If this option is switched off the peristaltic pump

will not be switched on at

Startup hardware

or at

the start of a determination.

Reference channel

Reference

channel.

On

If this option is switched on the absorption of

the measuring and reference channels will be

shown as a comparative value.

Wavelength, nm

Sets the wavelength of the reference channel.

Bunching, nm

Sets bunching for the wavelength defined

above.

Links

The

Links

tab in the window for the system settings serves to select and

set the COM port (for details see Software Instructions for Use «IC Net

2.3», Section 5.2.4 Links).

3.3.2 Hardware

settings

Selection of the

Hardware

menu item using the right mouse button

opens the hardware settings window consisting of the four tabs

Outputs

,

Alarm stops

,

System settings

and

Detector settings

.

Outputs

The

Outputs

tab of the hardware settings window defines remote output

signals to be set automatically if specific events occur.