5.1 Technical data

844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051

97

Immunity to interference

Standards met:

- EN/IEC 61326

- EN/IEC 61000-4-2

- EN/IEC 61000-4-3

- EN/IEC 61000-4-4

- EN/IEC 61000-4-5

- EN/IEC 61000-4-6

- EN/IEC 61000-4-11

- EN/IEC 61000-4-14

- NAMUR

5.1.14 Temperature control column heating

Temperature range

+10°C...75°C, controllable in 0.1 °C steps.
Only heating! The temperature cannot be

controlled to a temperature below the ambient

temperature.

Accuracy

± 0.5 °C

Stability

better than 0.05 °C

Scanning time

250 ms

5.1.15 Ambient

temperature

Nominal operating range

+5…+45°C

(at 20…80 % atmospheric humidity)

Storage

–20…+70°C

Transport

–40…+70°C

5.1.16 Housing

Material of cover

Aluminium

Material of base

Steel, enameled

Material of door

Polyurethane rigid foam (PUR) with fire protection

for fire class UL94VO, CFC-free

Width

259 mm

Height

381 mm

Depth

512 mm

Weight

ca. 25 kg