3.2.4 Instrument

icon

When you click with the right mouse button on the instrument icon, you
will see the following menu:

Open

Opens the window for

System Settings.

Hardware

Opens the window for

Hardware Settings.

Spectrum window

Opens the

SPECTRUM WINDOW

.

Unlink

Unlink device from the system.

3.2.5

Hardware start/stop and baseline recording

Start
Starting the hardware (by selecting

SYSTEM / Control / Startup

hardware

) on the 844 UV/VIS Compact IC includes starting the high-

pressure pump and, depending on the system start values

Startup with

hardware

defined on the

Configuration

tab, starting the UV lamp, the VIS

lamp, and, if present, the column heating and peristaltic pump.
At the same time Recording the measuring signal is started, using
the method defined in the system. The measuring signal is recorded
independently from the time for the chromatogram set under

Duration

,

until either data acquisition is stopped with

SYSTEM / Control / Stop data

acquisition

or a new determination is started. Alternatively data recording

can also be interrupted by clicking on in the chromatogram
window. In this case the user is asked whether the recorded baseline is

to be saved or not.

Stop
By selecting

SYSTEM / Control / Shutdown hardware

the high-pressure

pump, UV lamp, VIS lamp and, if present, the column heating and the
peristaltic pump of the 844 UV/VIS Compact IC will be stopped
immediately. The current determination and the processing of the active
sample table will also be stopped. In addition, the remote output lines
will be set to the values defined for the parameter

PowerON values

on

the tab

Outputs

in the

HARDWARE

window.

3.2.6 Determination

start/stop

Start
By selecting the menu item

SYSTEM

/

Control

/

Start determination

a

determination is started with the settings of the selected system. At the
start the system start values

Startup with hardware

defined for the system