2.3 Attaching the accessories

844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051

23

68

69

70

67

1 Mount connectors

Slide a pressure screw 69 (6.2620.000) and a ferrule 68

(6.2620.010) over the end of the capillary 70 to be fastened.

2 Insert capillary in connection

Push capillary end into the corresponding connection as far as it

will go (to avoid dead volume).

3 Tighten pressure screw

Tighten pressure screw 69 with the open-end spanner

1

/

4

"

(6.2621.010) supplied.

Figure 11: Connectors for Steel capillaries

2.3.5

Filter unit PEEK

With the versions 2.844.0010, 2.844.0020, 2.844.0110 and

2.844.0120, there is a 6.2821.120 PEEK filter unit mounted between

the high-pressure pump and injection valve

46. It protects the valve

against particles.
PEEK filter unit

34 consists of housing 73, filter screw 71 and filter 72.

New filters

72 are available as an option under ordering number

6.2821.130 (10 pieces).

When connecting the filter unit please note the flow direction arrow

printed on the housing.


67 Ferrule (6.2620.010)

used in pre-installed

components only

68 Ferrule (6.2620.270)

69 Pressure screw (6.2620.000)

70 Capillary

6.2620.020 Steel capillary