2.9 Detector
The 844 UV/VIS Compact IC is equipped with UV/VIS detector. The
eluent (and the sample) are led through a measuring cell (flow-through
cell
42/62) and measured spectroscopically. The available light
sources are a UV (deuterium) lamp and a VIS (halogen) lamp.
2.9.1
Installing the flow-through cell
With the 2.844.0030 und 2.844.0130, the flow-through cell is already
pre-installed!
Figure 25: Flow-through cell installation
42 6.2839.130 Flow-through cell PEEK
62 6.2839.100 Steel flow-through cell
129 Optical bank
130 Flow-through cell outlet "OUT"
Eluent outlet
131 Flow-through cell inlet "IN"
Eluent inlet
132 Knurled screws
For fastening flow-through cell 42/62.
To install flow-through cell
1 Remove blank cover
• Optical bank 129 is delivered with a blank cover. This should
be removed.
42/
62
130
132
131
129