4.2 Maintenance and servicing

844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051

87

2 Insert new pump tubing

• Insert the new pump tubing 94 in the tubing cartridge 96 as

shown in Figure 17. The orange-yellow stopper 95 must click

into the corresponding holder on the left-hand side of the

tubing cartridge.

• Place the tubing cartridge 96 on mounting pin 40 and press

down on the right-hand side until snap-action lever

98 clicks

into position on holding clamp 50. Take care that no kinks are

formed in the pump tubing.

3 Set contact pressure

• Switch on peristaltic pump.
• Press contact pressure lever 97 upwards until the solution

just starts to be drawn in. Then press contact pressure lever

upwards until it clicks once more to obtain optimal contact

pressure.

• Switch off peristaltic pump.

4.2.6

Replacing the UV lamp

Proceed as follows to replace the UV lamp:

The UV lamp becomes hot when used for a long time! Allow it to cool

down before you start to replace it.

Avoid contact with the surface of the lamp! Residues reduce the

light transmittance and may also be burned into the glass and cause

permanent damage.
If it becomes dirty the lamp surface should be cleaned with alcohol

before it is used again.

1 Insert lamp

Insert UV lamp 153 into the "Opening for UV lamp" 155 at the

bottom of lamp cooling housing 9.

2 Screw on fastening ring

• Screw down attachment ring 154. The attachment ring forms

part of the 844.

3 Connect cable

Plug "UV lamp connection cable" 11 into " UV lamp

connection"

12 of the 844 UV/VIS Compact IC.