2 connect the lamps, 3 flow-through cell tubing connections, 10 putting into operation – Metrohm 844 UV/VIS Compact IC User Manual

Page 61: Connect the lamps, Flow-through cell tubing connections

Advertising
background image

2.9 Detector

844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051

49

2 Attach flow-through cell

• Attach flow-through cell 42/62 to the optical bank with the

knurled screws 132.

2.9.2

Connect the lamps

Connect the UV (deuterium) and VIS (halogen) lamps as follows:

1 Connect VIS (halogen) lamp

• Plug "VIS lamp connection cable" 13 into "VIS lamp

connection"

10.

2 Connect UV (deuterium) lamp

Plug "UV lamp connection cable" 11 into "UV lamp

connection" 12.

2.9.3

Flow-through cell tubing connections

Make the flow-through tubing connections as follows:

1 Connect flow-through cell inlet

with 2.844.0010:

• Attach connection capillary 111 (6.1831.110) connected to

column 113 to inlet "IN" 131 of flow-through cell 42 using a

compression fitting

64 (6.2744.010).

with 2.844.0020:

• Attach connection capillary 107 (6.1831.110) to outlet "to

detector"

105 of post column reactor 101 using a

compression fitting 64.

Attach the other end of connection capillary 107

(6.1831.110) to inlet "IN"

131 of flow-through cell 42 using a

compression fitting 64 (6.2744.010).

mit 2.844.0030:

Connection capillary 112 (6.2620.120) connected to

column 113 is already connected to inlet „IN“ 131 of flow-

through cell 62 at delivery.

with 2.844.0110:

• Attach connection capillary 122 (6.1831.120) connected to

column 113 to inlet "IN" 131 of flow-through cell 42using a

compression fitting

64 (6.2744.010).

with 2.844.0120:

• Attach connection capillary 123 (6.1831.120) to outlet "to

detector"

105 of post column reactor 101 using a

compression fitting 64.

Attach the other end of connection capillary 123

(6.1831.120) to inlet "IN"

131 of flow-through cell 42 using a

compression fitting 64 (6.2744.010).

with 2.844.0130:

• Attach connection capillary 124 (6.1831.120) connected to

column 113 to inlet "IN" 131 of flow-through cell 62 using a

Advertising