2.6 High-pressure pump

844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051

31

2.6.3

Connecting the eluent bottle

The eluent supply line from the storage bottle to the high-pressure

pump is connected as follows:

Only

degassed (with N

2

, He or vacuum) and

microfiltered (0.45

μ

m

filter)

eluents should be used!

The 6.1608.070 Eluent bottle (2 L) supplied is

not suitable for vacu-

um degassing. Use a pressure-resistant container for this.

Care must be taken that the

eluent used is freely miscible with any

solvent remaining in the pump head (the pump head leaves the

factory filled with either isopropanol or methanol/water). If this is not

the case then the pump must first be rinsed with a solvent which is

miscible with both the previous eluent and the following eluent (e.g.

acetone).

1 Prepare eluent bottle

• Prepare, microfilter (0.45 μm microfilter) and degas (with N

2

,

He, or vacuum) the suitable eluent for the required

application and separating column.

• Fill eluent into eluent vessel 86(clear glass, 2 L).

2 Install bottle attachment

• Firmly screw threaded stopper 84 (6.1446.040; part of

6.1602.160) into the smaller threaded opening (M6) of bottle

attachment

85 (6.1602.105; part of 6.1602.160).

• Firmly screw aspirating filter 87 onto aspirating tubing 82.
• Pull the other end of aspirating tubing 82. through the larger

threaded opening (M8) of bottle attachment 85 from below.

• Push O-ring (E.301.0021; part of 6.1602.160) over the free

end of aspirating tubing

82. and move it towards bottle

attachment 85.

• Push tubing nipple 83 (4.420.4300; part of 6.1602.160) over

the free end of aspirating tubing

82., move it as far as requi-

red towards bottle attachment 85 and screw it loosely in the

larger opening of bottle attachment 85.

• Insert aspirating tubing 82. with screwed-on aspirating filter

87 into eluent bottle 86 and screw bottle attachment 85 onto

eluent bottle 86.

Pull aspirating tubing 82. so far through the opening of tubing

nipple 83 that aspirating filter 87 is touching the bottom of

eluent bottle 86.

Fix aspirating tubing 82. in place by screwing shut tubing

nipple 83.