New IC precolumns are normally filled with solution and sealed at both

ends. Before the precolumn is installed in the system, it must be

ensured that this solution is freely miscible with the eluent used

(check manufacturer's specifications).

1 Connect precolumn

• The precolumn is mounted upstream from the separating

column. Use as short a piece of 6.1831.010 PEEK capillary as

possible.

• If column heating is used (see Section 2.8.4 the precolumn

can also be placed in the inner compartment of the column

heating. Use an S-Guard for this. Other precolumns can only

be mounted outside the column heating.

When you install the column, always ensure that this is inserted

correctly in accordance with the flow direction (if existent) shown.

2 Rinse the precolumn

• Place a beaker beneath outlet capillary.
• Switch on IC Pump in «IC Net» and rinse precolumn for

approx. 10 min with eluent.

• Switch off IC Pump.

2.8.2 General

information

on separating columns

The precolumn that is suitable for your separating column can be

found in the

Metrohm IC Column Catalog, the product information

about separating columns that can be found under

http://www.met-

rohm.com

, ion chromatography products, or let your agency advise

you directly.

New IC separating columns are normally filled with solution and

sealed at both ends. Before the column is installed in the system, it

must be ensured that this solution is freely miscible with the eluent

used (check manufacturer's specifications).

The separating columns currently available from Metrohm and preco-

lumns can be found in the Metrohm IC Column Catalog, or on the

Internet under

http://www.metrohm.com

in the ion chromatography

product section. Each column is supplied with a test chromatogram

and a data sheet. More detailed information about special IC

applications can be found in the relevant "Application Bulletins" or

"Application Notes"; these are available on the Internet under

http://www.metrohm.com

in the applications sector, or can be

requested free of charge from your local Metrohm agency.

When you install the column always ensure that this is inserted

correctly in accordance with the flow direction shown.