5 replacing the pump tubing, Replacing the pump tubing, Figure 28 – Metrohm 844 UV/VIS Compact IC User Manual

Page 98: Components of inlet valve 141 and outlet valve 142, Er 151

Advertising
background image

Table of contents

Figure 28: Components of inlet valve

141

and outlet valve

142

4.2.5

Replacing the pump tubing

The pump tubings used by the peristaltic pump are consumable

material with a limited lifetime and should be exchanged at regular

intervals (approx. every 4 weeks under continuous use).
The working life of pump tubing depends to a considerable extent on

the contact pressure. This is why the contact pressure must be

correctly set as described in Section 2.10.2. If the pump is to remain

switched off for a lengthy period of time the tubing cartridges 96 should

be raised completely by loosening the snap-action lever

98 on the right-

hand side (the set contact pressure remains unchanged).
As the pump is always operated on the same side the 6.1826.110

Pump tubings supplied can be used on both sides. To exchange a

pump tubing proceed as follows:

1 Remove old pump tubing

Press contact pressure lever 97 on the tubing cartridge 96

down as far as it will go.

• Release tubing cartridge 96 from 50 by pressing down snap-

action lever 98 and remove from mounting pin 40.

• Remove old pump tubing 94.

146 Valve housing

147 Sealing ring (black)

148 Sleeve

149 Sapphire sleeve

The bright side must point towards the

sapphire sphere.

150 Sapphire sphere

151 Ceramic holder for sapphire

sphere

The large recess must point in the

direction of the sapphire sphere

152 Seal

The larger opening must point

outwards

146

147

148

149

150

151

152

146

147

148

151

150

149

152

141

Inlet valve

142

Outlet valve