Mains connection, 1 setting the mains voltage, Setting the mains voltage – Metrohm 844 UV/VIS Compact IC User Manual

Page 37: 78, filter screw 76 and filter 77, 77 are available as an option un, 4 mains connection

Advertising
background image

2.4 Mains connection

844 UV/VIS Compact IC / Gebrauchsanweisung 8.844.1051

25

74 Steel connection capillary

6.2620.020

from steel purge valve to steel filter

unit.

75 Steel filter unit 6.2620.180

76 Filter screw of filter unit

Part of 6.2620.180 Filter unit

77 Filter 6.2620.190

Part of 6.2620.180 Filter unit

78 Filter housing of filter unit

Part of 6.2620.180 Filter unit

79 Steel connection capillary

6.2620.020

from steel filter unit to steel pulsation

dampener

2.4 Mains

connection

Follow the instructions below for connecting to the power supply. If the

instrument is operated with a mains voltage set wrongly and/or wrong

mains fuse, there is a danger of fire!

2.4.1

Setting the mains voltage

Before switching on the 844 UV/VIS Compact IC for the first time, check

that the mains voltage set on the instrument matches the local mains

voltage. If this is not the case, you must reset the mains voltage on the

instrument as follows:

1 Disconnect mains cable

Disconnect mains cable from mains connection plug 21 of the

844 UV/VIS Compact IC.

2 Disconnect mains cable

Using a screwdriver, loosen fuse holder 22 below the mains

connection plug 21 and take out completely.

3 Check and change fuse if necessary

Carefully take the fuse installed for the desired mains voltage out

of fuse holder 22 and check its specifications (the position of the

fuse in the fuse holder is marked by the white arrow imprinted

next to the mains voltage range):

100120 V 1.6 A (slow-blow)

220240 V 0.8 A (slow-blow)

4 Insert fuse

Change fuse if necessary and reinsert in fuse holder 22.

5 Install fuse holder

Depending on the desired mains voltage, insert fuse holder 22

in the 844 UV/VIS Compact IC so that the corresponding mains

Advertising