Figure 19: Post column reactor

The post column reactor connections to the column and detector are

described in Section 2.8 and Section 2.9.

2.7.4 Reagent

bottle

connection

The reagent solution tubing from the supply bottle to the peristaltic

pump is mounted as follows:

1 Preparing the supply bottle for the reagent

• Fill the reagent solution into supply bottle 109 (amber glass,

1 L) and label the bottle.

Screw bottle attachment 108 onto supply bottle 109.

2 Connect aspirating tubing for reagent

100 Connection capillary

PTFE Capillary 6.1803.050,

length L = 20 cm

101 Post column reactor 6.2836.000

103 Connection capillary

PEEK Capillary 6.1831.110,

length L = 30 cm

104 Post column reactor "eluent in"

Inlet from column

105 Post column reactor "to detector"

Outlet to detector

106 Post column reactor "reagent in"

Inlet for post column reagent from

peristaltic pump

107 Connection capillary

PEEK Capillary 6.1831.110,

length L = 30 cm

If a column heating is used, the

capillary

124 is connected here.

104

105 106

103

107

100

64

101

101