Connection of the 813 compact autosampler – Metrohm 844 UV/VIS Compact IC User Manual

Page 68

Advertising
background image

Table of contents

5 Setting up the 844/838 system

• Select

File / Open / System

in the main window of the «IC Net »

software. Open the system for the 844 UV/VIS Compact IC

that you have created in Section 2.5.4.

• Add the 838 Advanced Sample Processor to the system (via

Setup / New devices / Install new devices..

, see «IC Net»

Instructions for Use, Section 4.4.4):

6 Creating a program for the 838

• Open the window with the system settings of the 838 (with a

double-click on the 838 icon or via the right-hand mouse key

open

).

• Click on the

Program

tab. The 838 Advanced Sample Pro-

cessor is controlled by a command sequence on the

Program

tab.

• Create a program (see «IC Net» Instructions for Use, Section

6.26.3):

At the command Scan Wait1 the 838 will wait for an

incoming Remote signal on line 3 (defined in the subtab

Confi-

guration / Scan

). It will receive this Remote signal from a Remote

command included in the time program of the 844 (see below).

The 844 and 838 are synchronized in this way.

For more detailed information about the control of the 838

Advanced Sample Processors via «IC Net» please refer to «IC