8 cal l 0 0, 8 cal l f 0, 88 cal 10 – Rice Lake FC-i/Si Series, A&D Weighing User Manual

Page 34

Advertising
background image

display the desired calibration weight.
(For example: Using 10kg calibration weight
that actually weighs 10.002kg.)

3. Press the ENTER key.

The calibration weight stops blinking
and “

Cal 0” appears.

4. Making sure that there is nothing on, or touching

the weighing pan, press the ENTER key.
When zero calibration is completed, the display
will show “

Cal f”.

If you don’t need span calibration, press the
CAL switch to exit from the calibration
procedure.

5. Place the calibration weight on the weighing pan

and press the ENTER

key.

When span calibration is completed, the display
will show “

&DOLEUDWLRQ(QG

”.

If the calibration weight is not what it should be,
an error will be displayed. Check if the weight is
correct and try again.

6. Press the CAL switch and re-attach the

calibration switch cover.
(End of the calibration procedure.)

ENTER















.

8

Cal

l

0

0





SFV

& D O L E U D W L R Q  = H U R 

Cala0

& D O L E U D W L R Q  Z H L J K W  N J 

1234567890123456789012

10.000















.

8

Cal

l

f

0





SFV

& D O L E U D W L R Q  6 S D Q 

Calaf

& D O L E U D W L R Q  Z H L J K W  N J 

1234567890123456789012

10.000

& D O L E U D W L R Q  = H U R 

Cala0

& D O L E U D W L R Q  Z H L J K W  N J 

1234567890123456789

--------

& D O L E U D W L R Q  6 S D Q 

Calaf

& D O L E U D W L R Q  Z H L J K W  N J 

1234567890123456789

--------















.

88

Cal

10





SFV

& D O L E U D W L R Q  ( Q G 

Cal

a

& D O L E U D W L R Q  Z H L J K W  N J 

123456789012345678901210.000

& D O L E U D W L R Q  6 S D Q 

Calaf

& D O L E U D W L R Q  Z H L J K W  N J 

1234567890123456789

err

If the scale will be moved to another place, set the gravity
acceleration value before calibration. The value must be of the
area where the calibration is to be done.

Advertising